BLOOD GAS ANALYSIS Model: EasyBloodGas

Firm / country of manufacture: MEDICA / USA

Quality standards: ISO, FDA, GMP

(Do not use gas cylinders)

STANDARD CONFIGURATION

  • The main unit comes with standard accessories;
1. Sample reagent (Reagent Module) 01 bottle
2. Reference electrode 01 piece
3. pH electrode 01 piece
4. The pO2 electrode 01 piece
5. PCO2 polarity 01 piece
6. Washing water 01 box
7. Thermal printing paper 02 rolls
8. Power cable 01 strand
9. Documentation of use 01 book

 

With the EasyBloodGas blood gas machine, Medica redefined the design of the blood gas machine

 

Introduce:

  • The sophistication of the traditional blood gas machine has been encapsulated in a sturdy design with a small stand to save space. The cylinders are removed. The gas and liquid have been mixed (using an barometer) to create a handy Reagent Module containing calibre solutions. All components are combined into three simple, easily accessible modules. for users. The periodic warranty is limited to the simple replacement of electrodes and pump lines.
  • Medica EasyBloodGas blood gas analyzer can measure pH, PCO2, PO2 and calculate other 11 parameters. Patient parameters include, Temperature, FIO2Hb can be added with an electronic keyboard and incorporated into the patient results. The measured and calculated results will be displayed on the screen and printed. Simple menus guide the user through the operation. The modular design will make maintenance easy.
  • EasyBloodGas addresses the economical need of the laboratory in returning patient results. Unique electrode design, combined with precise control of calibre volume ensure savings in operation and low cost per sample.

 

  1. Advantages:
  • Operation with the menu displayed simply by touching the keyboard:
  • EasyBloodGas can be programmed to follow predefined laboratory procedures. The software allows to select the required options including Reference Limits, Quality Control Limits, Operator ID and Patient Information.
  • Valve Module contains a very durable ceramic valve, selects and incubates the reagent
  • The Sample Module contains a multi-function sample aspirator and self-cleaning needle that is convenient and safe
  • Maintenance-free electrode design makes mounting fast and safe.
  • ... Easy to maintain:
  • EasyBloodGas can be maintained by anyone, anywhere and anytime:
  • The innovative design simplifies maintenance, addressing the need of the laboratory with limited access to technical staff. Technical support can be carried out by phone or fax, completely eliminating the need for on-site repair. Analysis software displays the condition of the parts for quick diagnosis. Modular design makes disassembly quick and easy.
  • 3 modules can be disassembled - Reagent, Valve, Sensor without any tools.
  • No need to use gas cylinders, bringing great benefits to users
  • The Reagent module independently removes the gas cylinders and stores all waste.
  • ... Easy to manage data:
  • General quality control and data management:
  • The EasyBloodGas quality control program calculates and stores all 3 levels of the last 30 quality test results. A printed Levey-Jennings chart identifies trends. The data management program compares the entire results of all patients in an array stored in memory and flag the results that exceed the threshold. Results for up to 64 patients can be stored in memory.
  • …Easy to use:
  • Manipulating blood gas analysis has never been simpler
  • The versatile aspirator can accept aspirated samples by syringe or capillary tube. The ability to automatically clean the suction needle creates convenience, keep the prototype and is safe for the user
  • The reagent module definitely makes it convenient, economical and safe
  • The bulky gas cylinders are replaced with compressed liquid calibres, all packed in a handy reagent module that can contain waste, thus protecting the user from the risk of infection. The reagent module's solid state memory allows EasyBloodGas to track the date and reagent usage. Ensure uninterrupted operation.
  • The electrodes are disposable and require no maintenance
  • Advanced membrane technology and novel fabrication techniques offer unmatched convenience in changing electrodes. Users never need to change membranes with Medica's integral membrane design, thus saving time and simplifying maintenance.

 

II. Technical information:

  • CLIA classification: Moderate complexity
  • Sample type: Whole blood
  • Sample capacity: 100ul Syringe, 75ul Capillary Tube
  • Measured parameters:
  • pH: 6.9 - 7.9 pH units
  • PCO2: 8.0 - 150.0 mmHg
  • PO2: 10 - 700 mmHg

 

  • Calculated parameters:
  • pH (T): The correctly calculated pH temperature
  • PO2 (T): PO temperature2 calculated correctly
  • PCO2 (T): PCO temperature2 calculated correctly
  • TCO2 (Total Carbon Dioxide): 0 - 50 mmol / L
  • HCO3– (Bicarbonate): 0 - 50 mmol / L
  • BEIGEb (Base Excess in Blood): -25.0 - 2.50 mmol / L
  • BEIGEecf (Base Excess in extra cellular fluid): -25.0 - 25.0 mmol / L
  • SBC (Standard Bicarbonate): 0 - 50 mmol / L
  • %SO2c (Oxygen Saturation calculated at normal P50): 40.0 - 100.0%
  • A - aDO2 (Alveolar arterial oxygen gradient): 0 - 700 mmHg
  • RI (Respiratory Index): 0.0 - 70.0

 

  • Input parameters:
  • Patient temperature: 5 - 45oC
  • Hemoglobin: 30 - 300 g / dL
  • FIO2: 10 - 100%
  • Patient ID: 14 digits

 

  • Sample temperature control: 37oC ± 0.2oC
  • Room conditions:
  • 15 - 30oC (59 - 86oF), 500 - 800 mmHg (up to 15 PSI)
  • Relative humidity 5 - 85%, no dust
  • Atmospheric environment (21% PO2)

 

  • Analysis time: 125 seconds
  • Data saving:
  • Results of 64 patients with Operator ID, Patient ID, date and time
  • 30 results per level (1, 2, 3)

 

  • Calib: Automatically or on request
  • Import Export: Numeric keypad, display screen, 27-column thermal printer, barcode reader port, RS-232 computer interface port

 

  • Electrical Power: 100/115 ~ VAC, 50 - 60 Hz, 0.8A or 220 ~ VAC, 50 - 60 Hz, 0.4A
  • Dimensions and mass:
  • 5 "W x 12.5" H x 7 "D (37 cm W x 32cm H x 18cm D)
  • 16 lbs (7.3 kg) including the reagent module

 

SPECIFICALLY NO NEED TO USE GAS BAG

en_USEnglish