MYTHIC 18 - AUTOMATIC HEMICAL TESTER 18 PARAMETERS, 3 PLASTIC PARTS APPLYING THE RESISTANCE PRINCIPLE

I. Principle

The machine performs tests on the principle of common impedance but has distinct applications

 The principle has not changed

This technology is based on modifying the impedance of a normalized slit immersed in an electrolyte and passed through a constant current generated by two electrodes placed on either side of the gap. A vacuum medium is created on one side of the slit that allows the cells to pass through. The physical volume of these cells is inversely proportional to the current flowing through it. An electrical pulse is recorded at the output of the electrode. The height of this electric pulse corresponds to the volume of the cell

The impedance principle is used to measure white blood cells, red blood cells, and platelets
   Differential application:

+ Constant current application of two electrodes

New system of counters, can be removed without the need for any tools.

+ The fluid flow system of the device is simplified, using a distribution pipeline system, reducing the length of the pipeline in the machine, minimizing the risk of pipe clogging.

+ New innovations about the sample suction module

+ New Invention of Pump Module

 The machine belongs to a new generation of equipment

+ A main circuit board does not need manual voltage measurement

+ Have software to adjust white tube of Hemoglobin

+ 24-V external power supply

+ The electrode valve does not need maintenance.

 

II. Specifications

 

Measured parameters 18 parameters and 3 charts

WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT, PDW.

LYM % and #, MON % and #, GRA % and #

 

Identification

 

- Identification by letters and numbers

Patient identification

- String type recognition

- Barcode reader (optional extra)

 

Wattage  

60 samples / hour

 

Sample volume  

9.8 µl

 

Chemical  

3 flasks (diluent, red blood cell breaking solution, washing solution)

 

Language  

French, English, Spanish, German, Italian, Portuguese, Chinese, Cyrillic ...

 

Electricity supply  

90-250V / 50-60Hz / <50VA

 

Size  

350 x 250x 340mm (H x W x D)

 

Friendly interface  

QVGA touch screen, numeric keypad built in the machine

 

Warning - Signaling abnormal results

- Signaling the Lab limits

- Chemical warning

- Warning equipment

 

CHeck quality  

6 lot control, 100 runs / lot, the LJ chart

 

Data saving  

Stores 1500 patient files, more expandable

 

Track record - System status

- Chemical control

- Standard mode

- QC (Quality Control)

- Operation service (repair)

Software (optional extra)  

 

- Software SVM: Application for generation Mythic machines

 

 

 Standards and quality control

Contact your local Orphee representative for information on the standards and quality control specified for each device.

en_USEnglish